Screening for intracranial hypertension using ambient-light-corrected computational pupillometry: a prospective pilot study
Article information
Abstract
Background
Intracranial hypertension is a life-threatening condition. Invasive intracranial pressure (ICP) monitoring remains the reference standard; however, catheter-based monitoring is not available for all at-risk intensive care unit patients. We evaluated whether ambient-light-corrected computational pupillometry could provide a bedside rule-out signal using the Pupil Reactivity (PuRe) score.
Methods
In this prospective single-center pilot study (2025–2026), adults who underwent intraparenchymal ICP monitoring underwent serial computational pupillometry (PuRe Pupillometer, Solvemed Inc.). The PuRe score is a 0–5 measure of pupillary reactivity (0, non-reactive; 0.1–2.9, sluggish; and ≥3, brisk). The primary outcome was diagnostic performance for ICP ≥22 mm Hg at the clinical brisk-reactivity boundary (PuRe score <3), using analyses that accounted for repeated measurements within patients.
Results
Twenty-six patients were included. At PuRe score <3, negative predictive value was 97.1%, sensitivity 81.1%, specificity 64.5%, and area under the receiver operating characteristic curve 0.77 (95% CI, 0.55–0.92). Mean ICP fell from 30.7 mm Hg with unreactive pupils to 7.6 mm Hg with brisk reactivity, and the inverse PuRe–ICP association held in a mixed-effects model accounting for clustering (slope –1.41 mm Hg per unit, P<0.001). The PuRe score showed no detectable dependence on ambient lighting in the ICP- and patient-adjusted models (P=0.13), whereas the constriction-based metrics remained light-dependent (P<0.001).
Conclusion
In this small single-center pilot cohort, preserved PuRe reactivity was infrequently accompanied by concurrent ICP ≥22 mm Hg. These preliminary findings support the further validation of ambient-light-corrected computational pupillometry as a potential adjunctive bedside screening, but not as a replacement for invasive monitoring.
INTRODUCTION
Intracranial hypertension is a bedside emergency requiring prompt recognition by neurocritical-care clinicians. It can complicate traumatic brain injury, subarachnoid hemorrhage, intracerebral hemorrhage, ischemic stroke with edema, hydrocephalus, and other acute brain injuries. As the intracranial pressure (ICP) increases, cerebral perfusion can decrease; sustained or untreated elevation may lead to tissue hypoxia, herniation, and death. Therefore, timely recognition is central to neurocritical care [1,2].
Invasive ICP monitoring using intraparenchymal probes or ventricular catheters remains the reference standard for quantifying the intracranial dynamics and guiding treatment. However, invasive monitoring requires neurosurgical expertise and specialized equipment and carries risks, including hemorrhage, infection, and device malfunction [3,4]. In practice, not every patient at risk of intracranial hypertension receives invasive monitoring. Many patients are monitored clinically until deterioration prompts imaging or catheter placement. In these cases, an earlier warning signal could be useful.
Therefore, non-invasive neuromonitoring is most clinically useful as a screen, not as a replacement for invasive monitoring. A reassuring non-invasive result can lower concern, whereas a concerning or deteriorating result can prompt repeat examinations, imaging, or invasive confirmation. This screen-then-confirm logic has been evaluated for transcranial Doppler and other non-invasive approaches [5-9]. Pupillary reactivity fits naturally into this framework because the pupil is already part of bedside neurological examinations.
However, manual penlight assessment is subjective and poorly suited for detecting subtle changes over time [10,11]. Quantitative infrared pupillometry has made pupil assessment more objective and has linked depressed reactivity to intracranial hypertension and poor outcomes [12-20]. A consistent pattern across this literature is clinically important; preserved reactivity is more useful for ruling out dangerous ICP elevation than depressed reactivity is for proving it [17]. Thus, pupillometry is best understood as a bedside rule-out signal.
However, ambient light remains a practical problem in serial pupillometry. The intensive care unit (ICU) lighting changes across night shifts, procedures, transportation, emergency evaluations, and operating-room conditions. If a reactivity measure changes with the room light, clinicians cannot confidently interpret the trend. Computational pupillometry addresses this problem by combining camera-based acquisition, controlled light stimulation, ambient-light sensing, and a reactivity score designed to remain interpretable across routine bedside illumination [21-25]. Camera-based pupillometry on consumer devices has shown feasibility and diagnostic utility in the context of brain injury, although its performance depends on careful design [26-28] and is attractive in settings where fixed infrared hardware is impractical [29,30]. Fig. 1 summarizes this framework, from the invasive reference standard to the bedside and ambient-light-corrected reactivity score.
Conceptual framework for computational pupillometry as a bedside screening tool for intracranial hypertension. (A) Invasive intraparenchymal intracranial pressure monitoring, the reference standard used in this cohort. (B) Computational pupillometry as a non-invasive bedside assessment of pupillary reactivity. (C) Mobile implementation of computational pupillometry, combining controlled light stimulation, ambient-light sensing, and on-device generation of the ambient-light-corrected Pupil Reactivity (PuRe) score. PLR, pupillary light reflex; NPU, neural processing unit; GPU, graphics processing unit.
The Pupil Reactivity (PuRe) score reports pupillary reactivity on a 0–5 scale: 0 indicates non-reactive pupils, 0.1–2.9 sluggish reactivity, and ≥3 brisk reactivity [24,25,31]. In this study, we evaluated ambient-light-corrected PuRe scores against invasive intraparenchymal ICP monitoring in adults with acute intracranial pathologies. The primary objective was to test whether the clinical brisk-reactivity boundary (PuRe score <3) could identify intracranial hypertension, defined as ICP ≥22 mm Hg, with emphasis on negative predictive value. The secondary objective was to characterize the relationships among PuRe reactivity, ICP, cerebral perfusion pressure, and ambient lighting under routine ICU conditions.
METHODS
Study design and participants
This prospective, single-center pilot study was conducted in a tertiary mixed medical-surgical ICU between 2025 and 2026. Because patients had impaired consciousness or were receiving ongoing analgosedation and could not provide consent, the bioethics committee granted a waiver of consent for this observational study. The reporting followed the STROBE guidelines [32].
Eligible patients were adults aged at least 18 years, admitted with acute intracranial pathology, and enrolled within 24 hours of intensive-care admission. Inclusion diagnoses were traumatic brain injury with Glasgow Coma Scale score ≤12, ischemic stroke with respiratory failure, intracerebral hemorrhage with National Institutes of Health Stroke Scale score ≥5, or subarachnoid hemorrhage with Hunt-Hess grade >1. Patients with a history of major ocular surgery, retinal disease, ocular prostheses, or acute brainstem pathology affecting pupillary reactivity were excluded from the study. Invasive ICP monitoring was performed when clinically indicated and ICP-related analyses were restricted to pupillometric recordings with ICP values documented at the same observation time.
Computational pupillometry and ICP monitoring
Pupillary examinations were performed using a computational pupillometry platform (PuRe Pupillometer, Solvemed Inc.) operating on a consumer smartphone. Each acquisition recorded the pupillary light reflex and, when image quality was sufficient, produced the PuRe score, an ambient-light-corrected 0–5 measure of pupillary reactivity: 0 indicates non-reactive pupils, 0.1–2.9 sluggish reactivity, and ≥3 brisk reactivity [24]. Standard pupillary-light-reflex metrics were also derived for comparison, including the initial pupil diameter, constriction amplitude, percent constriction, and maximum constriction velocity. These constriction-based metrics are the uncorrected pupillary parameters from which the ambient-light-corrected PuRe score is computed; therefore, comparisons between them isolate the contribution of the ambient-light-correction step. Recordings were obtained with the phone held approximately 15 cm from the eye using a 1-second baseline, 1-second light stimulus, and 3-second recovery. Ambient illumination was measured by the device and spanned approximately 10–1,000 lux, which is consistent with routine ICU lighting.
The ICP was monitored using an intraparenchymal microsensor (Codman Microsensor metal skull-bolt kit, REF 626638; Integra LifeSciences) placed by a neurosurgeon under sterile conditions. ICP values were documented in the electronic medical records, and each pupillometry recording was paired with the ICP value documented at the same observation time. The mean arterial pressure was recorded, and the cerebral perfusion pressure was calculated as the mean arterial pressure minus the ICP. Intracranial hypertension was defined using the guideline-aligned Brain Trauma Foundation treatment threshold of ICP ≥22 mm Hg [1,33]; an ICP ≥20 mm Hg threshold was additionally examined as a sensitivity analysis.
Statistical analysis
The analysis was organized according to four clinical questions. First, we asked whether the PuRe score differed between observations with and without intracranial hypertension. Second, we evaluated the diagnostic performance of the clinical brisk-reactivity boundary (PuRe score <3) for ICP ≥22 mm Hg, emphasizing negative predictive value, sensitivity, specificity, area under the receiver operating characteristic curve, and likelihood ratios [34]. Third, we tested whether the PuRe score changed with the ambient light and compared it with standard constriction-based metrics. Fourth, we examined whether repeated measurements in the same patient affected the interpretation of findings.
Continuous variables were summarized as medians (interquartile range) or means (standard error). PuRe reactivity categories were defined as non-reactive, sluggish, and brisk, and group differences were compared using a one-way analysis of variance. Because each patient contributed repeated bedside observations, recordings were not treated as independent: the continuous PuRe–ICP relationship was estimated with a linear mixed-effects model with a random intercept for patient, the binary outcome (ICP ≥22 mm Hg) with cluster-robust (patient) logistic regression and a population-averaged generalized estimating equation, and all diagnostic-performance confidence intervals by patient-level bootstrap; performance was also examined at the more inclusive ICP ≥20 mm Hg threshold. Diagnostic performance was evaluated at the pre-specified clinical brisk-reactivity boundary (PuRe score <3), with prevalence-independent likelihood ratios and a prevalence projection of predictive values. The lighting analysis evaluated the ambient illumination while accounting for repeated observations and concurrent ICP, complemented by two one-sided equivalence tests. Drug-level medication data were not systematically recorded; a structured search of the recording notes and metadata for sedative, opioid, and vasoactive drug terms returned no matches; therefore, a medication-adjusted analysis was not possible. The brisk-reactivity boundary (PuRe score <3) was the single pre-specified primary operating point; the equivalence margins (±0.3 and ±0.5 PuRe units, both narrower than the one-unit width of the reactivity categories) and all other thresholds, models, lighting analyses, and outcome comparisons were exploratory and hypothesis-generating rather than confirmatory. Because this boundary was fixed in advance rather than fitted to the data, out-of-fold threshold selection was not required. Analyses were performed in Python 3.13 (NumPy, pandas, SciPy, statsmodels, scikit-learn) and interpreted as exploratory because this was a single-center pilot study.
RESULTS
Study cohort
Twenty-six consecutive patients met the eligibility criteria, and their characteristics are summarized in Table 1. The cohort was 69% male, with a median age of 52 years (interquartile range, 43–62) and a median admission Glasgow Coma Scale score of 5 (interquartile range, 4–13). The primary diagnoses were subarachnoid hemorrhage in 12 patients, traumatic brain injury in 10, ischemic stroke in 2, and no structured diagnosis code available for 2 patients who met the eligibility criteria for chart review. Patients were monitored throughout their intensive-care stay, yielding 1,342 pupillometry recordings with a median of 48 per patient (interquartile range, 43–66). ICP-related analyses used the pupillometry recordings that had a contemporaneously documented ICP value, because each ICP value was entered together with its recording rather than carrying a separate observation timestamp; the pupillometry and ICP were contemporaneous by design and no time-matching window was required. Because the cohort was selected for invasive ICP monitoring, it had high-acuity by design. Intracranial hypertension, defined as ICP ≥22 mm Hg, was uncommon, occurring in 9.2% of paired observations.
Single-patient bedside trajectory
A single-patient bedside trajectory illustrates the pattern that the cohort-level analysis tests (Fig. 2). Over the monitoring period, the PuRe score decreased as the ICP increased, with concordant changes in mean arterial pressure and within-session anisocoria. This trajectory is illustrative and provides face validity at the bedside; however, it is not used as evidence of cohort-level diagnostic performance.
Bedside trajectory from a representative neurocritical-care patient. (A) Intracranial pressure (ICP) over 39 hours of monitoring, with the 22-mm Hg threshold marked. (B) Pupil Reactivity (PuRe) score by eye over time, with brisk, sluggish, and non-reactive zones shaded. (C) Initial pupil diameter (INIT) by eye. (D) Mean arterial pressure (MAP). (E) Within-session anisocoria of INIT. (F) PuRe score against concurrent ICP. The trajectory is illustrative and shows the bedside pattern of falling reactivity accompanying rising ICP; it is not used as evidence of cohort-level diagnostic performance.
Pupillary reactivity, ICP, and perfusion
Across the cohort, the average pupillograms showed a progressively dampened light reflex as the ICP increased, with less reactive pupils accompanying intracranial hypertension (Fig. 3). The relationship was graded across the three reactivity categories, from brisk to unreactive: mean ICP rose from 7.6 mm Hg with brisk reactivity to 9.9 mm Hg with sluggish reactivity and 30.7 mm Hg with non-reactive pupils (Fig. 4C); this across-category gradient was descriptive (one-way analysis of variance F=99.3, P<0.001, treating observations as independent). The formal, clustering-aware analysis was a linear mixed-effects model with a random intercept for patient, which accounts for repeated measurements: each one-unit increase in PuRe score corresponded to a 1.41 mm Hg decrease in ICP (95% CI, –2.09 to –0.74; P<0.001; residual intraclass correlation, 0.88). A within/between decomposition separated a within-patient slope of –1.30 mm Hg per unit (95% CI, –1.98 to –0.62) from a larger between-patient slope of –7.37 mm Hg per unit (95% CI, –11.72 to –3.02), the within-patient slope being roughly one fifth of the between-patient slope. This indicates that the categorical ICP gradient across reactivity strata was driven predominantly by between-patient differences rather than by the PuRe score tracking ICP changes within individual patients over time; therefore, within-patient tracking was limited, and the PuRe score was best interpreted as a rule-out risk marker rather than as a continuous within-patient ICP surrogate. Cerebral perfusion pressure showed the opposite gradient, rising from 57 mm Hg with non-reactive pupils to 80 mm Hg with brisk reactivity.
Pupillary light reflex by intracranial pressure (ICP) category. (A, B) Averaged pupillograms for recordings with ICP <22 mm Hg and ICP ≥22 mm Hg; the shaded band marks the light stimulus. Pupils recorded during intracranial hypertension had larger baseline diameter and less constriction. (C-F) Mean pupillometry parameters across the same ICP categories: initial pupil diameter (INIT; C), constriction amplitude (CAMP; D), percent constriction (DELTA; E), and the Pupil Reactivity (PuRe) score (F). All four parameters differed between strata in descriptive across-category comparisons (one-way analysis of variance, P<0.001 each); the formal inference accounts for within-patient clustering in the main analysis.
Ambient-light stability, intracranial pressure (ICP) association, diagnostic performance, and outcome association of the Pupil Reactivity (PuRe) score. Bars show mean (standard error). (A) PuRe score in dark and bright recordings by ICP category. (B) Constriction amplitude (CAMP), percent constriction (DELTA), and maximum constriction velocity (MCV) across the same lighting categories, standardized to a shared axis. (C) Mean ICP by PuRe reactivity category, falling from 30.7 mm Hg in non-reactive observations to 7.6 mm Hg in brisk observations (descriptive across-category comparison; one-way analysis of variance F=99.3, P<0.001); the formal clustering-aware inference is the mixed-effects model reported in the text. (D) Mean ICP in patients who died versus survived (mean and standard error); descriptive only. (E) Pupillary metrics by the same patient outcome, standardized to a shared axis; panels (D) and (E) are descriptive associations, not causal attributions, and the outcome comparison is underpowered. (F) Receiver operating characteristic curve for the PuRe score discriminating ICP ≥22 mm Hg, with AUC 0.77 (95% CI, 0.55–0.92) and the clinical brisk-reactivity boundary marked. (G) Lighting-induced change in each metric from dark to bright recordings, expressed in units of that metric’s standard deviation; a smaller value indicates greater lighting stability. (H) Modeled dark-to-bright counterfactual: the estimated proportion of observations whose intracranial-hypertension classification would change between dark and bright ambient conditions (a modeled estimate from the fitted lighting slope, not directly observed paired flips); the PuRe score shows the smallest shift. INIT, initial pupil diameter; AUC, area under the receiver operating characteristic curve.
Diagnostic performance for intracranial hypertension
At the clinical brisk-reactivity boundary (PuRe score <3), preserved reactivity was rarely accompanied by intracranial hypertension. Negative predictive value was 97.1% (95% CI, 93.2–99.7), sensitivity was 81.1%, specificity was 64.5%, and the area under the ROC curve was 0.77 (95% CI, 0.55–0.92) (Table 2, Fig. 4F). The boundary operates at the level of the individual reading rather than as a one-time patient-level clearance, and it was fallible: 10 observations in 4 patients showed ICP ≥22 mm Hg despite a brisk PuRe score (false negatives), consistent with the 81.1% sensitivity and with serial, rather than single, assessment. In a univariate logistic-regression classification of ICP ≥22 mm Hg with cluster-robust (patient) standard errors, each one-unit increase in PuRe score lowered the odds of intracranial hypertension (odds ratio, 0.49; 95% CI, 0.28–0.83, P=0.009); a more conservative population-averaged generalized estimating equation was directionally consistent but did not reach significance (odds ratio, 0.73; 95% CI, 0.40–1.35, P=0.32). Because predictive values depend on disease frequency, we also report prevalence-independent likelihood ratios: the negative likelihood ratio was 0.29 (95% CI, 0.05–0.76) and the positive likelihood ratio was 2.28 (1.31–4.60); projected negative predictive value ranged from 98.5% at 5% prevalence to 83.7% at 40% prevalence (positive predictive value 10.7% to 60.4%). The positive predictive value in this cohort was 18.9% (95% CI, 6.8–37.9). Performance was similar at the more inclusive ICP ≥20 mm Hg threshold (negative predictive value, 95.7%; 95% CI, 90.8–99.3; sensitivity, 77.3%; specificity, 65.2%), indicating the rule-out signal is not an artifact of the chosen threshold. Therefore, the PuRe score should be read as a rule-out signal rather than as a confirmatory test.
Stability across ambient lighting
Across the naturally variable lighting of the ICU, in a repeated-measures model adjusted for ICP and within-patient clustering, the PuRe score showed no detectable dependence on ambient illumination (P=0.13), whereas the constriction amplitude, percent constriction, and maximum constriction velocity remained light-dependent (P<0.001 each) (Fig. 4A and B). At the screening threshold, a modeled counterfactual dark-to-bright shift changed the classification of 1.4% of PuRe-scored observations compared to 5.8% to 7.3% for the constriction metrics (Fig. 4G and H); this decision-scale comparison was an exploratory illustration rather than an independent validation of ambient-light stability. Two one-sided equivalence tests did not certify equivalence at the pre-specified ±0.3 and ±0.5 PuRe-unit margins, and bright-condition observations were comparatively sparse, so formal equivalence was not established. These findings illustrate why ambient light can be a clinically relevant confounder for conventional pupillary-light-reflex metrics and why ambient-light-corrected scoring is important for serial ICU measurements.
Outcome association
The vital status at discharge was available for 25 patients; 13 died and 12 survived. Recordings from patients who died showed higher mean ICP than those from survivors (14.6 vs. 7.0 mm Hg) and more depressed reactivity across pupillary metrics (Fig. 4D and E). With only 25 patients, these outcome comparisons were underpowered and reported descriptively because the cause of death was not recorded, whether mortality reflected neurological deterioration or systemic causes could not be determined, and these results are associations rather than attributions.
DISCUSSION
In this single-center pilot cohort, ambient-light-corrected computational pupillometry produced a PuRe score that behaved as a bedside rule-out signal for intracranial hypertension. At the clinical brisk-reactivity boundary, preserved reactivity was rarely accompanied by ICP ≥22 mm Hg, while absent or sluggish reactivity was associated with substantially higher ICP. The score also showed no detectable dependence on the ambient lighting, unlike standard constriction-based metrics. Because this cohort was already undergoing invasive monitoring, the findings support the rule-out signal itself; the role of frequent non-invasive surveillance before invasive monitoring in patients not yet monitored is a hypothesis for a prospective study rather than the use demonstrated here.
The pupil is not a pressure transducer. ICP can change rapidly with positioning, ventilation, sedation, cerebrospinal-fluid dynamics, osmotherapy, and surgical or procedural interventions. The pupillary light reflex reflects an integrated brainstem and autonomic response and can be affected by sedatives, opioids, ocular disease, and baseline pupil characteristics. Therefore, a modest association between an individual pupil measurement and the simultaneous ICP value is expected. A more relevant clinical question is whether the preserved reactivity makes dangerous ICP elevations unlikely. In this cohort, the categorical gradient was clear: mean ICP was 30.7 mm Hg in non-reactive observations and 7.6 mm Hg in brisk observations. This separation was driven mainly by between-patient differences; the within-patient slope was modest and roughly five-fold smaller than the between-patient slope. Therefore, the score is better suited to identifying lower- or higher-risk readings or clinical states than to tracking moment-to-moment ICP changes within a patient.
This pattern is consistent with prior quantitative-pupillometry literature, where preserved reactivity has tended to be more useful for ruling out intracranial hypertension than depressed reactivity has been for ruling it in [15-20]. In practice, a low PuRe score should not be used by itself to diagnose intracranial hypertension. This should prompt the clinician to examine the patient, review the trend, consider confounders such as medications or ocular disease, and escalate to imaging or invasive monitoring when clinically indicated. Conversely, a persistently brisk PuRe score may provide bedside reassurance when concordant with other neurological examinations. However, because the sensitivity and negative-likelihood-ratio confidence intervals were wide, and the population-averaged generalized estimating equation did not reach significance, this pilot should be interpreted as compatible with, rather than definitive evidence of, rule-out utility.
These findings motivated the prospective evaluation of a two-tier neuromonitoring model. Invasive ICP monitoring remains the reference standard and is required when pressure must be measured and treated directly. Computational pupillometry could in principle serve earlier in the pathway, providing frequent non-invasive surveillance for patients who are at risk but not yet invasively monitored; because the present cohort was already monitored, this earlier role is a prospective hypothesis rather than a finding of this study. Patients with depressed or deteriorating PuRe reactivity, new asymmetry, or discordant clinical features should be escalated to imaging and invasive monitoring, where ICP can be confirmed and quantified. This framing makes computational pupillometry complementary rather than competitive to invasive monitoring.
Ambient light is not only a technical nuisance but also a practical ICU problem. Pupillary measurements may be obtained during the night shift, after lights are turned on, during procedures, during transport, or in bright emergency and operating rooms. Serial interpretation becomes difficult if pupillary metrics change with room illumination. In this study, the PuRe score showed no detectable dependence on ambient lighting, whereas the constriction amplitude, percent constriction, and maximum constriction velocity remained light-dependent. At the decision level, the lighting-related classification shift was smaller for the PuRe score than for the constriction metrics. Formal statistical equivalence across lighting was not certified in this small pilot study; therefore, we describe the PuRe score as designed for ambient-light correction and substantially more stable across lighting than constriction-based metrics, rather than as proven equivalent across illumination. Because the PuRe score is a machine-learning composite derived from ambient-light-corrected pupillary parameters, whereas the comparators are the uncorrected forms of the same inputs, this contrast is best understood as an engineering validation of the ambient-light-correction step rather than as an independent clinical comparison, demonstrating that clinical superiority over an existing dedicated infrared pupillometry platform would require a separate head-to-head study. A paired-device study has reported agreement between a phone-based computational pupillometry system and a dedicated infrared pupillometer for pupillary-light-reflex parameters [35]. The present study adds an evaluation against the invasive ICP reference standard and supports the central design rationale of computational pupillometry: a reactivity score should remain interpretable across routine bedside illumination.
This study has several important limitations. This was a single-center pilot study involving 26 patients, which limited its precision and generalizability. The cohort was predominantly male and selected for invasive ICP monitoring; therefore, it represents a high-acuity population and may not be generalizable to the emergency department, ward, transport, or pre-hospital settings. Intracranial hypertension is uncommon; therefore, sensitivity, specificity, and likelihood-ratio estimates are imprecise; the sensitivity CI spanned 48.7 to 97.2% and the negative likelihood ratio CI extended to 0.76, a range that encompasses both adequate and inadequate rule-out performance. The most conservative clustering-aware analysis, a population-averaged generalized estimating equation, did not show statistical significance for the primary screening question (odds ratio, 0.73; P=0.32). These wide intervals leave open the possibility that rule-out performance is insufficient, and the projected negative predictive values carry the same imprecision as the underlying likelihood ratios; the negative predictive value will also depend on local disease prevalence and require confirmation in a larger sample. Sedatives, opioids, vasoactive medications, and ophthalmic comorbidities can affect pupillary dynamics and were not captured at the level of individual drugs and doses. Sedation is a bidirectional confounder; analgosedation can depress pupillary reactivity independently of ICP, which would bias toward over-calling elevated ICP. However, sedatives also directly lower ICP, and during sedation lightening, ICP may rise before pupillary reactivity recovers, potentially resulting in false negatives, a scenario most damaging to a rule-out test. Therefore, the net direction of this confounding factor could not be determined without drug and dose data. The cause of death was not recorded; therefore, the outcome differences were associated rather than attributable. Pupillometry timing and frequency followed clinical care rather than a fixed schedule; therefore, the monitoring cadence varied across patients and could influence within- versus between-patient decomposition. Larger multicenter studies with prospective threshold validation, standardized measurement timings, medication-aware data collection, and patient-level false-negative assessments are required before clinical use.
This single-center pilot study offers an early clinical demonstration that ambient-light-corrected computational pupillometry may support rule-out screening for intracranial hypertension. A brisk PuRe score was rarely accompanied by ICP ≥22 mm Hg, while non-reactive pupils accompanied substantially higher ICP and, in descriptive comparisons, lower cerebral perfusion pressure. The PuRe score also showed no detectable dependence on ambient lighting, unlike the standard constriction-based metrics.
These findings suggest that computational pupillometry is a potential bedside component of two-tier neuromonitoring: frequent non-invasive surveillance before invasive confirmation. A larger multicenter validation is the next step in validating computational pupillometry as a screening method for elevated ICP.
Notes
Ethics statement
The study protocol adhered to the Declaration of Helsinki and was approved by the Independent Bioethics Committee for Scientific Research at the Medical University of Gdansk (No. KB/240/2025). Informed consent was waived because patients had impaired consciousness or were receiving ongoing analgosedation and could not provide consent.
Conflict of interest
Several authors are employees or equity holders of the company that developed the computational pupillometry application evaluated in this study and are co-inventors of related issued and pending patents. No other potential conflicts of interest relevant to this article were reported.
Funding
None.
Acknowledgments
The authors thank the nursing and medical staff of the intensive care unit at the study center for their support during data collection.
Author contributions
Conceptualization: MW, RC, MAW, KC. Methodology: MW, RC, MDz. Software: JC, MS, HC, MDz, RC. Formal analysis: RC, MW. Investigation: KC, MAW, AF, RMR, GG, JK, PK, SZ, MJK, PW, MSz, ML, KPS. Resources: MAW, KC. Data curation: RMR, JC, MS. Supervision: MAW, KC, MW. Writing - original draft: KC, MW, RC. Writing - review & editing: KC, MAW, AF, RMR, JC, MS, GG, JK, PK, HC, MDz, MW, RC, SZ, MJK, PW, MSz, ML, KPS. All authors read and agreed to the published version of the manuscript.